High Cap Case Study

Possible Loads for the Same 2.00” balance

NF (lb)

PM (in-lb)

SF (lb)

YM (in-lb)

RM (in-lb)

AF (lb)

@ 0 PM

@ 0 NF

@ 0 YM

@ 0 SF

 

 

7,000

21,000

3,000

9,000

5,000

500

5,000

14,000

700

4,000

5,400

500

6,400

20,000

2,000

6,000

6,000

600

3,000

9,000

1,500

3,000

8,000

900

 

 

Calibration Results

The ABCS calibration data is used to determine the balance accuracy, all of the original loading data is back calculated using AIAA 6x96 matrix coefficients.

Standard Deviations % of Balance Maximum Design Load

COMPONENT

N1

N2

Y1

Y2

R1

X1

Accuracy at high roll load

0.02%

0.03%

0.04%

0.06%

0.14%*

0.03%

Accuracy at mid roll load

0.02%

0.03%

0.03%

0.05%

0.05%

0.03%

Accuracy at low roll load

0.02%

0.03%

0.02%

0.03%

0.03%

0.02%

* The error on roll is usually a bit higher than the other components due to design trades in optimizing the load capacities and output of all of the components.